Defibrillator -AED-BeneHeart S2


Physical specifications
Dimensions: 210 mm (W) x 286 mm (L) x 78 mm (H)
Weight: 2.3 ± 0.3 kg (with screen version, including 1 battery)
Display 1
Screen type: color LCD
ECG display: no
Working mode: automatic, indoor and outdoor (automatic mode requires automatic selection of screen brightness according to ambient light brightness)
Size: 7 inches
Resolution: 800 × 480 pixels
Dustproof and waterproof grade: IP55
Temperature: working environment: - 5 ~ 50 ° C
Storage environment: - 30 ~ 70 ° C
Humidity: working / storage environment: 5 ~ 95% (non condensing)
Atmospheric pressure: working / storage environment: 57.0 ~ 106.2 kPa
Impact: rtca-do-160g-2010, section
7Operational Shocks;
IEC 60601-1-12,10.1.3,10.1.4
Vibration: mil-std-810g-2008, method
514.6,Category 13,Category
14,Category 20,Category 24;
EN13718-1,4.7.2;
IEC 60601-1-12,10.1.3,10.1.4
Collision: en1789, 6.3.4.2; EN13718-1,4.7.2
Drop: 1.5 m (once for each of 6 surfaces)
EMC:YY 0505; GB9706.8
Battery
Battery type: disposable lithium manganese battery
Battery capacity: standard battery: 4200 Mah
High capacity battery: 6600 MAH (optional)
working hours
Screen version: working time shall not be less than 12 hours; 350 200 J discharges (new battery, charge and discharge 3 times per minute)
Standby time
Standard battery: 5 years (ambient temperature 20 ± 5 ° C, daily self inspection, no power on, no self inspection report)
High capacity battery: 8 years (ambient temperature 20 ± 5 ° C, daily self inspection, no power on, no self inspection report)
Low battery reminder: at least 30 minutes of working time and at least 10 200 J defibrillation discharges (ambient temperature 20 ± 5 ° C)
Electrode sheet specification
Applicable population: Adults / children
Cable length: 1.2 m, pre connected
Validity period: 5 years after delivery
Equipment self inspection
Self inspection frequency: daily / weekly, monthly, quarterly
Equipment status indicator: visually display the current equipment status
Data storage
Event data: up to 500
Recording data: up to 1 hour
CPR data: up to 5 hours
Self inspection report: 1000 copies
Data export: support USB data export
Electrode piece: not universal
Validity period of electrode sheet: 5 years

STARS:
Catalogs for more options